The American Society of Electroneurodiagnostic Technologists website contains lots of interesting and useful information about skills of technologist, for example 10-20% system EEG electrodes placement, described below.

10-20% system
10-20% system
10-20% system
10-20% system

Tools of the Trade

As with any profession, a tech is only as good as the tools used. You will need to have marking pencils in two different colors, a centimeter tape measure, calipers and hair clips (for patients with long hair).

The marking pencils need to be sharp to avoid thick marks which may contribute to errors. Why measure in millimeters if you make marks that are 5 mm wide?

Our patient model
Our patient model

Inion-Nasion measure

Measure Inion-Nasion distance
Measure Inion-Nasion distance
Divide Inion-Nasion distance on two to get central line (CZ)
Divide Inion-Nasion distance on two to get central line (CZ)

Doing your math

Measure from nasion to inion. In this case the total is 35 cm.

Take 10% of that number (3.5 cm) and make a mark 3.5 cm above the nasion (FpZ position).

Take 20% of the 35 (7 cm) and add that to the 3.5. Make a mark at 10.5 cm (FZ), then 17.5 (CZ), and 24.5 (PZ).

Your final mark should be 10% (3.5 cm) above the inion mark (OZ).

Continuing sagittal measurements

You have now made the marks that will determine the anterior-posterior positions of Fpz, Fz, Cz, Pz and Oz. (although Fpz and Oz are not usually used in routine work.

Pre-auricular to Pre-auricular

Measure pre-auricular to pre-auricular distance
Measure pre-auricular to pre-auricular distance

Section the hair from ear to ear going through the vertex. Ask the patient to open and close their mouth. Find the indentation created by the temporo-mandibular joint when the mouth is open and mark that point on each side. † Measure between those marks going through the vertex (Cz) mark made in last step. In our case this distance is 35 cm.

Pre-auriculars continued

Take 10 % of the total (3.5 cm) and make a mark above the pre-auricular mark you created. Then take 20% of the total (7 cm). Add 7cm to 3.5 and make a mark at 10.5. Continue adding 7 cm to the previous number: then mark 17.5 cm and 24.5 cm. The last mark should be 3.5 cm from the mark over the ear.

Alternate Method Hint: †An alternate method is to just move the tape each time, always lining up the 0 with the previous mark and marking at 7. Some say this can contribute to mistakes but so can adding numbers incorrectly!

First Position Determined

At this point, you will have a cross mark at the intersection of the vertex. This is the CZ location. As measurement continues, you will make the cross marks at every electrode position.

Circumference

Measure circumference
Measure circumference

Section hair around the circumference using the Fpz, Oz and marks just above the ears as a guide.

Now using the tape as a guide, place it under the Fpz mark and the mark above the ear. Make a mark above the tape along the top edge of the tape at the approximate location of F7, T5, T6 and F8.

In our case this distance is 55 cm.

Measure circumference
Measure circumference

Circumference continued

Measure the total circumference (55 cm). Take 10% of the total (5.5 cm) and set the calipers to 5.5 cm. Place the caliper legs with the nose evenly dividing them and make a mark at the tip of each leg. The final mark should be 5.5 cm from the first mark.† You should now have cross marks at F7, T3, T5, T6, O1, O2, T6, T4 and F8.

Using the FpZ mark as a guide for the level, make the cross marks at Fp1 and Fp2.

Using the FpZ mark as a guide for the level, make the cross marks at Fp1 and Fp2
Using the FpZ mark as a guide for the level, make the cross marks at Fp1 and Fp2

Parasagittals

Parasagittals
Parasagittals

Section the hair from Fp1 to O1 going through the mark at the central electrode position. Measure the total from Fp1 to O1. Total is 24 cm. Take 25% of total which is 6 cm and make a mark at 6 cm, 12 cm and 18 cm. Mark above the tape.† You should now have a cross mark at the C3 position. Repeat for the right side being sure to mark above the tape just like you did on the left.

Parasagittals
Parasagittals

POSTERIOR TRANSVERSE

Section hair from T5 to T6 going through the marks at the parietals.†Measure from T5 to T6. Total is 24 cm. Divide 24 cm by 25%. Make a mark at 6 cm, 12 cm, and 18 cm. You now should have cross marks at T5, P3, Pz, P4 and T6.

TRANSVERSE FRONTAL

†Section hair from F7 through Fz to F8. Measure the to distance from F7 to F8.† Total is 22 cm. Take 25% of that which is 5.5 cm. Make a mark at 5.5 cm, one at 11 cm and one at 16.5 cm. There should be cross marks now at F7, F3, FZ, F4 and F8.

Parasagittals
Parasagittals

SECTIONING HAIR

As the final step, section the hair and secure with hair clips or rubber bands. When the hair is all clipped out of the way, electrode application can be achieved much more efficiently.

SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
SECTIONING HAIR
Prepare skin on the marked points
Prepare skin on the marked points

Final Inspection

Look at the marks to see if they appear to be anatomically correct. During the Measurement and Application section of the board, now is a good time to remeasure and check electrode positioning.

Are they anatomically correct? Did you mark nasion and inion so the examiners can easily identify them?

T1 and T2 Measurments

Measure from pre-auricular point to the outer canthus of the eye.† Make a vertical mark at one third the distance. Make a mark along the top edge of the tape. Measure up one centimeter from the last mark and intersect with the vertical mark.

T1 and T2 Measurments
T1 and T2 Measurments
Electrodes placement
Electrodes placement
Electrodes placement
Electrodes placement
Electrodes placement
Electrodes placement
Electrodes fixation
Electrodes fixation
Electrodes fixation
Electrodes fixation

Software helpful tools

Modern EEG software contains special tools, called EEG electrodes locator. which can help technologist to put EEG electrodes according 10-20% system.

You need just measure and set to the software three sizes:

  • Inion-Nasion
  • Pre-auricular to Pre-auricular
  • Circumference
Software will calculate all distances automatically
Software will calculate all distances automatically

And the software will calculate distances between each EEG electrode automatically.

This function is most convenient during impedance measuring mode during electrodes placement:

During impedance measure software shows all distances between electrodes
During impedance measure software shows all distances between electrodes

Useful materials

  1. EEG Long-Term Monitoring Guidelines (ILAE and IFCN).
  2. QEEG. Quantitative EEG tools in modern clinical practice.
  3. EEG in ICU.
  4. Neurosoft Neuromonitor system.
  5. Personalized MDD treatment with TMS (Individual Alpha Peak Frequency).
  6. Applying EEG to Multimodality Neurologic Monitoring in Pediatric Neurocritical Care.